Input
 
Type
Strain gauge, isolated pressure sensor
Range (psi)
 
100, 200, 300 (contact Telog
for other ranges)
Over pressure (psi)
300, 600, 600
Burst pressure (psi)
850, 1000, 1000
Resolution
0.025% of full scale, 12-bit
Accuracy
 
±0.25% of full scale at constant temperature
Temperature effect
±0.01% of full scale per ºC
Recording
 
Sample rate
Programmable from 4/sec up to 8 hours
Data interval
Programmable from 1/sec up to 8 hours
Values saved
Selectable min, avg & max per interval
Memory
128 Kbytes (~80,000 data values)
Interface
 
Type
RS-232; 300 to 19.2 Kbaud
Connector:
Circular 4 pin watertight
Battery
 
Type
 
Single AA Lithium (Saft LS 14500 or equal)
 
Life
five years typical
 
 
Physical
 
Temperature Ranges
 
  Operating
-10º C to +65º C
  Storage
-40º C to +65º C
Humidity
0-100% relative humidity
Enclosure
NEMA 4x/IEC IP65
Size
12.70 cm diameter x 8.89 cm
(5” diameter x 3.5”)
Thread
 
 
Hydrant mount: 2.50” NHT standard
Contact Telog for non-standard thread
Internal mount:
1/4 NPT
Support Software
 
S-3PC
Telogers for Windows
S-3PCL
Telogers for Windows Lite
Data Transfer Unit
 
Palm Pilot with
Telog support software
Computer Requirements
 
For S-3 PC or S-3PCL
IBM compatible computer with a 586/133 MHz or higher processor running on:
Microsoft Windows 95/98/NT/ 2000/XP. At least 32 MB of RAM, a hard disk with at least 200 MB of free space and a pointing device.
 

Copyright

Rep Website